Instrumented Charpy Pendulum Impact(withdrawn 2-19-2020)
Test ID
M-428
Description
A Dynatup POE 2000 apparatus is used to impact a specimen. Specimens are bars with dimensions specified in ASTM and notched accordingly. Specimen is placed horizontally, with notch centered opposite of impact side. Machine is connected to a computer to generate a load deflection curve. At a set velocity and energy, pendulum is released to impact specimen. Test is complete when pendulum completes one swing. Five replicates are performed.
Options
Standards
- ASTM D6110
- ISO 179
Data
Break Type, Impact Strength. This data appears in Matereality as Charpy Impact.
Replicates
Number of times this test is repeated: 5
Samples
10 notched Charpy specimens [details]
Application
This test is used as a means of quality control or materials specification. It can, in some cases, be used on a finished product to determine if the part will fracture under the set conditions. Data gathered from this test are valuable for comparative purposes, to aid the engineer in choosing an appropriate material for a specific application.